    摘要:

    为提高解离增强镧系荧光免疫分析(DELFIA)的灵敏度或信/噪比,进行了一些重要的方法学研究.观察到了对于不同的铕量,荧光响应和信/噪比都随着增强液体积而明显地变化.对于确定的铕量,存在一个最佳体积,且铕量越小,其最佳体积也越小.实验中选择最佳体积是重要的.研究了增强液制备技术,发展了微滴定板条有效的清洗和干燥方法,使本底荧光明显降低.

    Abstract:

    Factors influencing the sensitivity, or the signal/noise ratio.of dissociation enhanced lanthaqide fluoroimmunoassay (DELFIA) have been studied. The fluorescence responses and signal/noise ratios for different europium amount were shown to be changed with the volume of enhancement solution.and there was an optimum volume at a certain europium amount.The smallest europium amount leads to the smallest optimum volume.20/00 of the net fluorescence intensity was increased by using tinfoil reflection layer. Effective washing and drying methods of microtitration strips decreased background fluorescence have been developed.

    参考文献
    相似文献
    引证文献
引用本文

赵启仁,张福华,刘洁,林汉.解离增强镧系元素荧光免疫分析灵敏度的改进[J].生物化学与生物物理进展,1994,21(3):255-258

复制
分享
文章指标
  • 点击次数:
  • 下载次数:
  • HTML阅读次数:
  • 引用次数:
历史
  • 收稿日期:1993-05-03
  • 最后修改日期:1993-08-02
  • 接受日期:
  • 在线发布日期:
  • 出版日期: