使用ZS-3型半自动生化分析仪对血清丙氨酸氨基转移酶,总蛋白,肌酐等项目的实验室分析表明,该仪器的精密度试验重复性良好, 其批内CV 5%以内, 批间CV 9%以下;准确度分析显示三项试验平均回收率在95%~105%之间; 交叉污染实验提示吸入量在500 μl以上时,其交叉污染率可控制在2%~5%.
Clinical biochemical items such as alanine amino-transferase(ALT), Creatinine(Cr), total protein(TP) in serum were analyzed by using ZS-3 semi-automatic biochemical analyzer. The results show that the machine had a good precision. The mean intra-assay deviation CV was under 5% and mean inter-assay deviation CV was under 9%. The average recovery was 95%~105%. When the drawn liquid volume was over 500 μl, the cross contamination could be controled under 5%~2%.
李国君,沈文梅,王成彬. ZS-3型半自动生化分析仪的临床应用[J].生物化学与生物物理进展,1997,24(5):474-476