改良红细胞葡糖-6-磷酸脱氢酶活性直接测定法
DOI:
CSTR:
作者:
作者单位:

作者简介:

通讯作者:

中图分类号:

基金项目:


A Modified Assay of Glucose-6-Phosphate Dehydrogenase in Red Blood Cell
Author:
Affiliation:

Fund Project:

  • 摘要
  • |
  • 图/表
  • |
  • 访问统计
  • |
  • 参考文献
  • |
  • 相似文献
  • |
  • 引证文献
  • |
  • 资源附件
  • |
  • 文章评论
    摘要:

    建立一个在全自动生化分析仪上,去除6-磷酸葡糖酸脱氢酶(6PGD)对结果的影响且能准确、快速检测葡糖-6-磷酸脱氢酶(G6PD)活性的定量方法.采用自动扣减样本空白的速率法对107例正常人,31例G6PD明显缺陷病人(经高铁血红蛋白还原率法筛选)血样本进行测定.表明灵敏度达0.27 U/gHb.批内CV=5.6%;批间CV=9.4%.线性范围在0~15 U/gHb.与常规的高铁血红蛋白还原率法结果比较,两法相关系数(r)=0.863.速率法特异性较好,有结果稳定、准确、操作简便、检测快速等优点,是一种值得推广使用的方法.

    Abstract:

    To develop a precise and rapid method for glucose-6-phosphate dehydrogenase (G6PD) assay on the auto-analysis eliminating the interruption of 6-glucose-phosphate dehydrogenase (6GPD).Human blood samples from 107 normal people and 31 patients with obvious G6PD deficiency (Methemoglobin MHb) were analyzed by Rate A with sample blank correction is desired. Results showed that the measured sensitivity was 0.27 U/gHb, the within-run precision was 0.145. the coefficient of variation (CV) was 5.6% and the between-run precision was 0.229. the CV was 9.4%.The linear was between 0~15 U/gHb.Compared to the results measure by Rate A with sample blank correction is desired. The correlation coefficient(r)to two methods was 0.863.These results indicate that it is a sepificity, stable, simple and rapid method and this method is deserved to be popularized.

    参考文献
    相似文献
    引证文献
引用本文

关广雄,曹予明,沈雁.改良红细胞葡糖-6-磷酸脱氢酶活性直接测定法[J].生物化学与生物物理进展,2001,28(5):764-766

复制
分享
文章指标
  • 点击次数:
  • 下载次数:
  • HTML阅读次数:
  • 引用次数:
历史
  • 收稿日期:2000-09-11
  • 最后修改日期:2000-09-28
  • 接受日期:
  • 在线发布日期:
  • 出版日期: