Measurement of Creatine Kinase MM Isoforms by Chroiuatofocusing
DOI:
Author:
Affiliation:

Clc Number:

Fund Project:

  • Article
  • |
  • Figures
  • |
  • Metrics
  • |
  • Reference
  • |
  • Related
  • |
  • Cited by
  • |
  • Materials
  • |
  • Comments
    Abstract:

    A rapid and sensitive method of isoforms of CK (EC 2.7.3.2) -MM in human serum by chromatofocusing was reported. The assay system involved Mono P (HR5/20) column. fast protein liquid chromatography (FPLC) system and post column reactor. The result of CK-MM isoforms was obtained within 40 min after serum application to equilibrated column.The lower limit of sensitivity for CK activity is 8U/L and the linear capacity is 17-2085U/L, with CV values of 2.1%. 2.7% and 11.0% for MM1. MM2 and MM3 respectively.

    Reference
    Related
    Cited by
Get Citation

Kong Qingyin, Yang Zhenhua, Yang Shude, Tang Zhiyi. Measurement of Creatine Kinase MM Isoforms by Chroiuatofocusing[J]. Progress in Biochemistry and Biophysics,1994,21(1):83-85

Copy
Share
Article Metrics
  • Abstract:
  • PDF:
  • HTML:
  • Cited by:
History
  • Received:January 16,1993
  • Revised:April 20,1993
  • Accepted:
  • Online:
  • Published: