Determination of Several Cholesterol Autoxidanon Products by High Performance Liquid Chromatography
DOI:
Author:
Affiliation:

Clc Number:

Fund Project:

  • Article
  • |
  • Figures
  • |
  • Metrics
  • |
  • Reference
  • |
  • Related
  • |
  • Cited by
  • |
  • Materials
  • |
  • Comments
    Abstract:

    A sensitive and precise method for simultaneous determination of some major cholesterol autoxidation products(COPs)was described. The method was based on high performance liquid chromatography (HPLC) with 6-chlorostigmasterol as internal standard.After saponification and extraction, the COPs were derivatized into benzoates for measurement by HPLC with ultraviolet detection. Each product gave a linear response(r>0.9990) over a concentration range of 0.312~10.000 mg/L. The intra-and inter-assay co-efficients of variation was 2.15%~5.40% and 1.23%~3 .70%, respectively. This method was successfully applied to the determination of COPs in cholesterol standard reference material and in human serum.

    Reference
    Related
    Cited by
Get Citation

Dong Jun, Chen Wenxiang, Li Jianzhai. Determination of Several Cholesterol Autoxidanon Products by High Performance Liquid Chromatography[J]. Progress in Biochemistry and Biophysics,1996,23(2):179-182

Copy
Share
Article Metrics
  • Abstract:
  • PDF:
  • HTML:
  • Cited by:
History
  • Received:May 05,1995
  • Revised:July 28,1995
  • Accepted:
  • Online:
  • Published: