By utilizing imported chemiluminescent reagent, self-prepared rabbit anti-human type Ⅳ collagen IgG and a horseradish peroxidase-labeled anti-human type Ⅳ collagen monoclonal antibody, a method of sandwich chemiluminescent enzymimmunoassay(CLEIA) was established then evaluated. The results showed the CLEIA has a lower detecting limit of 30 μg/L,the average recovery was 94.5%, CVs of intra-assay and inter-assay were 6.0% and 11.6% respectively, the CLEIA method has a good correlation with Japan kits (ELISA, r=0.9686). the serum reference level by testing 60 cases health is 160 μg/L. It was found that the CLEIA is a sensitive, acurate, stable and specific method,it may be a qualified candidate for serum type Ⅳ collagen detemination of non-radioactive method.
WANG Tian-Cheng, YAO Min-Jie, LIU Xiang-Yi, KOU Li-Jun. Measurement of Serum Type Ⅳ Collagen by Sandwich Chemiluminescent Enzymimmunoassy[J]. Progress in Biochemistry and Biophysics,1998,25(6):554-556