Preparation of Triglyceride Enzyme Sensor and Its Application
DOI:
Author:
Affiliation:

Clc Number:

Fund Project:

  • Article
  • |
  • Figures
  • |
  • Metrics
  • |
  • Reference
  • |
  • Related
  • |
  • Cited by
  • |
  • Materials
  • |
  • Comments
    Abstract:

    A triglyceride enzyme sensor has been prepared by immobilizing lipase on pH glass electrode. The immobilized methods of lipase and the response characteristics of the sensor were studied.The linear range of detection is 3.09×10-6~1.91×10-3 mol/L for triglyceride with a slope of 32.7 mV/pC in the pH 8.5 Tris-HCl buffer solution at 37℃. The response time is 5~10 min. The sensor is relatively stable for 25 days.The sensor has been applied to the determination of triglyceride in serum with satisfactory results.

    Reference
    Related
    Cited by
Get Citation

WANG Yong-Sheng, LI Gui-Rong, Lü Chang-Yin, WAN Zai-Yang, LIU Chuan-Xiang. Preparation of Triglyceride Enzyme Sensor and Its Application[J]. Progress in Biochemistry and Biophysics,1999,26(2):144-146